Browse Categories
Search

High Precision Tension / Compression Load Cell with Digital Indicator, Software & Laptop, 50 g to 1 kg Capacity - EQ-TGA-LC-LD


Please email for lead time

Product Options

Capacity:

EQ-TGA-LC-LD is a high precision tension / compression load cell (50, 100, 200, 500, 1000 g capacity) with digital indicator, software & laptop. It has high performance (0.05% error full scale), high safe overload (1000%), software control and data logging, and compatibility with vacuum and controlled gas atmosphere, which makes it an ideal tool for customized gravimetric analysis system, such as thermogravimetric analysis (TGA).


SPECIFICATIONS:
Model / Capacity
  • 50 g (0.5 N), 100 g (1 N), 200 g (2 N)
  • 500 g (5N), 1 kg (10 N)
Load Cell Accuracy
  • Nonlinearity: 0.05% full scale, max
  • Hystersis: 0.05% full scale, max
  • Nonrepeatability: 0.05% full scale, max
  • Creep, in 20 min: 0.05%, max 
  • Zero balance: 2% full scale, max
  • 0.05% full scale, max
  • 0.03% full scale, max
  • 0.02% full scale, max
  • 0.025%, max
  • 3% full scale, max
Load Cell Temperature
  • Compensated Temperature: -10 to 45 °C
  • Operating Temperature: -55 to 90 °C
  • Temperature Effect on Output: 0.002%/°C, max 
  • Temperature Effect on Zero: 0.004%/°C, max
  • -10 to 45 °C
  • -25 to 80 °C 
  • 0.002%/°C, max 
  • 0.003%/°C, max
Load Cell Safe Overload

  • Safe Axial Overload: 1000% of capacity
  • Safe Side Overload: 500% of capacity
  • 1000% of capacity
  • 500% of capacity
Load Cell Calibration
  • Tension (default) or compression calibration 
  • NIST certified calibration with traceable calibration file
  • Tension (default) or compression calibration 
  • NIST certified calibration with traceable calibration file
Load Cell Electrical
  • Excitation Voltage: 12 VDC max
  • Rated Output: 2 mV/V
  • Cable Length: 1.5 m
  • 15 VDC max
  • 2 mV/V
  • 1.5 m
Digital Indicator
  • Power: 100 - 240 V
  • Display Update: 5/sec
  • Reading Speed: 230 readings per second
  • Error: 0.01% of reading. max
  • Operating Temperature: -10 to 50 °C
Laptop & Software Control
  • Software for load cell / indicator setup, display, data logging and output is pre-installed on the laptop and fully configured 
                                                       
Size
  • Load Cell: 50 mm L x 21 mm W x 50 mm H
  • Indicator: 96 mm L x 130 mm W x 48 mm H
  • Load Cell: 60 mm L x 16 mm W x 63 mm H
  • Indicator: 96 mm L x 130 mm W x 48 mm H
Net Weight
  • 5 kg
Warranty
  • One year limited warranty with lifetime support
Operation Instructions
  • Operation manual coming soon

 

Related Products

Shopping Cart
Your cart is empty.
Please clear the browsing history before ordering the product. Otherwise, availability and price are not guaranteed.
Mailing Lists
MTI sponsors MTI-UCSD Battery Fabrication Lab:
MTI-UCSD Battery Fabrication Lab Opening


MTI sponsors MRS and ECS Postdoctoral Award:

MTI Sponsors the Postdoctoral Awards

Upcoming Shows:
BPS 2018 62nd Annual Meeting Biophysical Society Feb 17-21 San Francisco CA Booth 909

2018 SME Annual Conference & Expo Feb 25 - 28 Minneapolis, MN Booth 701

APS March Meeting 2018 Mar 5-9 Los Angeles CA Booth 123

TMS 2018 147th Annual Meeting Exhibition Mar 11-15 Phoenix AZ Booth 500

Recent Attended Shows:
2017 MRS Fall Meeting and Exhibit Nov 26 - Dec 1 Boston, MA

Click here to see our past conferences

 

Notices
1. Prices listed online are valid for US market and whom pay by credit card only. There will be extra charges for shipping & handling. Price various from country to country.
2. We may add extra charges for paper order and net30 terms due to extra labor cost.
3. The prices listed are subject to change without notice.