Browse Categories
Search

200ºC Max.Compact Tape Casting Coater with Heated Vacuum Bed ( 10"Wx16"L ) & Doctor Blade - MSK-AFA-H200A


Transported by LTL Freight (Truck)

Product Options

Voltage:

Personalization

Oillless Vacuum Pump OFP (+ USD$599.00)

MSK-AFA- H200A is an NRTL standard Compact Tape Casting Coater with the vacuum chuck hotplate up to 200ºC and the active carbon filter on the top cover to protect lab environment. It is an excellent tool to make high-temperature polymer coating or conductive ceramic film on various substrates.

SPECIFICATIONS:
Power
  • 208 - 240V AC 50/60Hz
  • 1500W 
Heatable Vacuum Platform

  • Vacuum chuck made of aluminum alloy with micro-size hole
  • Vacuum chuck Area:   250mm(W)  x 400 mm(L) ( 10"W x 16" L )
  • Heating elements are installed inside chuck which can reach 200ºC Max. 
  • Precision digital temperature controller with +/- 1ºC accuracy
Vacuum Pump (Optional)
Film Applicator


  • Built in pushing bar for using micrometer adjustable applicator.
  • 180 mm width Applicator with micrometer adjustable head is included for immediate use (Click the picture below to see details).
  • 24VDC motor driven the pushing bar continuously to the adjustable position, which can be used under Ar gas glovebox.
           Note:  Thinnest film can be achieved is > 10 micron   (Click here to order other size film applicators)

Max. Coating Area
  • 350mm(L) x 250mm(W)  
Accuracy of Film thickness
  • +/- 0.01mm
Traverse  Speed & Distance
  • Travel Speed:        0 - 80mm/sec variable
  • Travel distance:     0 - 400 mm  adjustable
Top Cover and Filter
  • Active carbon filter and  venting fan are installed on the top cover 
  • The filter will remove organic vapor to make lab environmental safe & clean
Application Notes

  • You need a vacuum mixer to remove air bubbles inside slurry and to make uniform coatings (Click Pic.1 for detail.)
  • You may need a Slurry Viscosity Tester to make coatings consistent (Click Pic.2 for detail.)
  • You may need a vacuum oven to dry the films and remove binder after coating. (Click Pic.3 for detail.)
  • Rolling pressing is necessary to ensure good density and adhesion of electrode materials to the substrate after coating. Please select a rolling press from here. (Click Pic.4 for detail.)
  •     Pic.1     c.2        Pic.3        Pic.4             
Machine Dimensions
Click the Pic to see details
Net weight 40 kg (90 lbs)
Compliance

  • CE certified
  • NRTL or CSA certification are available upon request at the exta Cos
Warranty One Year limited with lifetime support
Operation Video
&
Instructions
                 

Related Products

Shopping Cart
Your cart is empty.
Please clear the browsing history before ordering the product. Otherwise, availability and price are not guaranteed.
Mailing Lists
MTI sponsorships:
MTI Sponsors Thermoelectrics Workshop

MTI-UCSD Battery Fabrication Lab


MTI Sponsors the Postdoctoral Awards

Upcoming Shows:
MS&T 2018 Oct 14-18 Columbus OH Booth 513

2018 MRS Fall Meeting and Exhibit Nov 25 - 30 Boston MA Booth 901

APS March Meeting 2019 March 4-8 Boston MA

Recent Attended Shows:
Furnaces North America Oct 8-10 Indianapolis IN

AiMES 2018 Sept 30 - Oct 4 Cancun Mexico


Zhejiang University MSE Visiting August 10th Richmond CA


Click here to see our past conferences

 

Notices
1. Prices listed online are valid for US market and whom pay by credit card only. There will be extra charges for shipping & handling. Price various from country to country.
2. We may add extra charges for paper order and net30 terms due to extra labor cost.
3. The prices listed are subject to change without notice.