KSL-1500X-GS125 is a bottom loading larger box furnace ( 125 Liter capacity) with inert gas atmosphere control up to 1400 ºC. The furnace consists of high-quality alumina fiber bricks and SiC heating elements with 20" x 20" x 20" heating chamber. A vacuum-sealed steel chamber with a water cooling jacket, as well as the vacuum & gas flow control system enables material processing under inert gas or oxygen atmosphere.
SPECIFICATIONS
|
- Double-layer stainless steel structure for maximum safety
- The bottom loading mechanism by electric hydraulic lift , which can hold 200 Kg Max. sample
- Water cooling front door and sealing jacket for max. service life.
- Touch-screen digital control panel for all parameters.
- Precision temperature control with +/- 3 ºC uniformity inside the chamber and +/-0.5°C accuracy.
|
Inside Chamber Size
|
- Heated chamber: 500 x 500 x 500 mm ( ~ 20" x 25" x 25" . 125 Liter capacity )
- Energy-saving 1800 grade fibrous alumina insulation.
- High purity Al2O3 coating.
|
Power
|
- AC 220V, three-phase; 50/60 Hz
- 40 KW
- 50Amp ( 80Aair-breaker)
|
Heating Element
|
- 12 Pcs SiC heating elements
|
Continuous Working Temperature
|
|
Maximum Working Temperature
|
- 1400°C (< 1 hr)
- Optional: Max. 1700oC is available by using MoSi2 heating elements at the extra cost
|
Heating Rate
|
5°C / min
|
Temperature Accuracy
|
+/- 0.5 ºC
|
Thermocouple
|
Two S type thermocouples (control thermocouple and secondary alarm thermocouple)
|
Temperature Control

|
- Eurotherm controller with PID automatic control via SCR (Silicon Controlled Rectifier)
- 30 programmable segments for precise control of heating/cooling rate and dwell time.
- Over-temperature protection and thermocouple failure alarm functions allow operation without attendant(s).
- +/- 0.5 ºC temperature accuracy.
- RS485 Communications Port
|
Gas Flow Control and Vacuum Valve |
- One flow meter ( 0~12L/min) is installed on the front panel of the furnace.
- One heavy-duty vacuum pump (CE certified) is built on the bottom of the furnace.

|
Net Weight
|
|
Warranty
|
One year limited warranty with lifetime support. (Consumable parts such as sample plates and heating elements are not covered by the warranty, please order the replacement at related products below.) |
Compliance
|
- CE Certified for the furnace, vacuum pump, and water chiller.
- NRTL (UL61010) or CSA certification is available at extra cost.( please click mark below to learn details
-

|
Application Notes
|
|
Warnings
|
- ATTENTION: Any damages caused by the use of corrosive and acidic gasses are not under the coverage of MTI One Year Limited Warranty.
- In order to prevent oxidation/corrosion damages to the Mo heating elements, it is necessary to keep a constant inert gas (Ar preferred) flow throughout the entire process. This is the most effective way to carry out any by-products.
|
Operation Instructions
|
|
|
|
| | |
|