|
1400C Max. Bottom Loaded Inert Gas Atmosphere Box Furnace, (20x20x20", 125L) - KSL-1500X-GS125
|
Please email for lead time
Item Number: KSL-1500X-GS125
Transported by LTL Freight (Truck)
|
KSL-1500X-GS125 is a bottom-loading larger box furnace ( 125 Liter capacity) with inert gas atmosphere control up to 1400 ºC. The furnace consists of high-quality alumina fiber bricks and SiC heating elements with a 20" x 20" x 20" heating chamber. A vacuum-sealed steel chamber with a water-cooling jacket, as well as the vacuum & gas flow control system, enables material processing under an inert gas or oxygen atmosphere.
SPECIFICATIONS
|
- Double-layer stainless steel structure for maximum safety
- The bottom loading mechanism is by electric hydraulic lift which can hold 400 Kg Max. sample
- Water cooling front door and sealing jacket for max. service life.
- Touch-screen digital control panel for all parameters.
- Precision temperature control with +/- 3 ºC uniformity inside the chamber and +/-0.5°C accuracy.
|
Inside Chamber Size
|
- Heated chamber: 500 x 500 x 500 mm (~ 20" x 20" x 20", 125 Liter capacity)
- Energy-saving 1800-grade fibrous alumina insulation.
- High purity Al2O3 coating.
|
Bottom Loading Door
|
- The bottom door is driven by an electric driven hydraulic cylinder up to 5 metric T
- Max. sample loading is ~400 kg (including crucible)
|
Power
|
- AC 208-240V, 3p, 50/60 Hz;
- 40 KW
- 380V, 400V, 460V, and 480VAC are also available upon request.
|
Heating Element
|
- 12 Pcs SiC heating elements
|
Continuous Working Temperature
|
|
Maximum Working Temperature
|
- 1400°C (< 1 hr)
- Optional: Max. 1700oC is available by using MoSi2 heating elements at the extra cost
|
Heating Rate
|
5°C / min
|
Temperature Accuracy
|
+/- 0.5 ºC
|
Thermocouple
|
Two S type thermocouples (control thermocouple and secondary alarm thermocouple)
|
Temperature Control
|
- Eurotherm controller with PID automatic control via SCR (Silicon Controlled Rectifier)
- 30 programmable segments for precise control of heating/cooling rate and dwell time.
- Over-temperature protection and thermocouple failure alarm functions allow operation without attendant(s).
- +/- 0.5 ºC temperature accuracy.
- RS485 Communications Port
|
Gas Flow Control and Vacuum Valve |
- One flow meter ( 0~12L/min) is installed on the front panel of the furnace.
- One heavy-duty vacuum pump (CE certified) is built on the bottom of the furnace.
|
Net Weight
|
|
Warranty
|
One-year limited warranty with lifetime support. (Consumable parts such as sample plates and heating elements are not covered by the warranty, please order the replacement at related products below.) |
Compliance
|
- CE Certified for the furnace, vacuum pump, and water chiller.
- NRTL (UL61010) or CSA certification is available at extra cost.( please click mark below to learn details
-
|
Application Notes
|
|
Warnings
|
- ATTENTION: Any damages caused by the use of corrosive and acidic gasses are not under the coverage of MTI One Year Limited Warranty.
- In order to prevent oxidation/corrosion damages to the Mo heating elements, it is necessary to keep a constant inert gas (Ar preferred) flow throughout the entire process. This is the most effective way to carry out any by-products.
|
Operation Instructions
|
|
|
|
| | |
|
Your cart is empty.
Please clear the browsing history before ordering the product. Otherwise, availability and price are not guaranteed. | |
| | |
|
MTI sponsorships: MTI Sponsors Thermoelectrics Workshop
MTI-UCSD Battery Fabrication Lab
MTI Sponsors the Postdoctoral Awards
Upcoming Shows:
|
|