1100C Hydrogen Gas Tube Furnace with 5" Quartz Tube with H2 Detecting System - OTF-1200X-S5-H2
|
In stock
Item Number: OTF-1200X-S5-H2
Transported by LTL Freight (Truck)
|
OTF-1200X-S-5-H2 is a single zone hydrogen tube furnace for the heat treatment of oxidation-sensitive materials up to 1100 ℃. The 5” processing tube is made of high-quality fused silica. The hydrogen detecting system comes with a Honeywell hydrogen gas detector, two solenoid valves, and an alarm siren. In case H2 leak is detected, the special design automatically activates the siren, cuts off the hydrogen source, shuts down the power, and purges nitrogen. It can also avoid air backflow if a power outage happens. Flange fixtures, full quartz tube covers, and double-layer steel furnace casing are installed to provide additional safety protection in case of emergency. The furnace is also compatible with all kinds of inert gasses as well as oxygen. The precision temperature controller provides customers to set up their own programmable heating profiles with up to 30 temperature segments.
Specifications:
Furnace Features
|
- Hydrogen gas detecting system with leak and power outage protection.
- Full tube covers, flange support, and double-layer steel casing with air cooling for extra protection.
- The hydrogen gas furnace must be placed under a fume hood which is not included. Please click to view the spec of the MTI EQ-FH-36 fume hood.
|
Power |
- AC 208-240 Single Phase, 50/60 Hz
- 3.0 KW Max. (20A breaker required)
|
Working Temperature |
- 1000 °C for continuous operation.
- 1100°C (< 1 hour)
|
Max. Heating Rate |
|
Max. Cooling Rate |
|
Processing Tube |
click here to choose a crucible
|
Heating Zone

|
|
Temperature controller
|
- PID automatic control via solid-state relay with 30 steps programmable
- Built-in Over Temperature and Thermocouple Failure Protection
- Temperature Accuracy: +/ - 1°C
- Optional: The furnace may be operated by PC with installing a control module. Please order an MTS02-B control module at an extra cost
|
Flanges and Fittings Left Flange Right Flange
|
- 5" flanges with quick load-lock doors for fast sample loading and retrieving.
- The water-cooled flange is available at an extra cost.
- Left flange assembly:
Mechanical pressure gauge Needle and solenoid valves 1/4" thermocouple feedthrough Two 1/4" tube fittings
- Right flange assembly:
H2 torch 0-1000 ml floating flow meter Hydrogen torch Gas-outlet valve
- Full tube covers and flanges fixtures to avoid hydrogen hazards in case of emergency or explosion.
-

|
Hydrogen Detecting System
(included) |
- One Honeywell UL approved sense point gas detector is Integrated into the system.
- Working Temperature Range: -40 ºF to +149 ºF (-40ºC to +65ºC)
- Alarm Point: H2 Lower Explosive Limit (LEL) is set at 20%. Once H2 gas is detected beyond the limit, it will cut off the hydrogen source, turn off the heating elements, and purge nitrogen.
- Power outage protection: if the furnace encounters a power outage, hydrogen source and heating elements will be cut off, and start to flow nitrogen.
|
Flow Meter and Hydrogen Torch
|
- One 1000 ml/min floating flow meter is installed on the gas-outlet end to adjust the flow rate.
- 1/4" stainless steel tube is installed as hydrogen torch to burn out the exhaust.
- One lighter is included in the standard package for igniting hydrogen gas manually.
|
Dimension
|
- Main furnace with lid closed: 440mm(L) × 440mm(D) × 460mm(H)
- Main furnace with lid open : 440mm(L) × 440mm(D) × 700mm(H)
- With flange support (260mm each), total: 960mm(L) × 650mm(D) × 700mm(H)
- Temperature Controller Unit: 260mm(L) × 350mm(D) × 410mm(H)
|
Net Weight |
|
Shipping Weight |
|
Shipping Dimension |
|
Warranty |
- One year limited warranty for electrical and mechanical parts
- Attention: A free training of safety operation at MTI is required to purchase this furnace. Training at the customer's facility is also available upon request at an extra cost. If users believe that they have sufficient experience and knowledge to handle the furnace in the dangerous or emergent situation, a liability release agreement must be signed at the purchase. MTI Corp. is not responsible for any damages caused by misuse without safety operation training. Any damages caused by the corrosive and acidic gases are not under the coverage of the MTI One Year Limited Warranty.
|
Compliance |
|
Warning
 Click here to learn the installation of a gas regulator
|
- Tube furnaces with quartz tubes are designed for use under vacuum and low pressure < 0.12 atm (absolute pressure)
- The flow rate for gases should be limited to < 200 SCCM for reducing thermal shocks to the tube
- Attention: A two-stage pressure regulator must be installed on gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn the installation of a gas regulator.
- Vacuum limit for all quartz tube furnaces: * furnace operation under vacuum is up to 1000 °C
|
|
|
| | |
|