OTF-1600X-III- is a split-able three-zone tube furnace with Kanthal 1650°C grade SiC heating element, 82mm OD Alumina tube and vacuum-sealed flange. Three 30 segments of temperature controllers are installed with S type thermal couples. Microprocessor-based self-tuning PID provides excellent control with no overshooting and  +/- 1°C accuracy. It is an excellent furnace for exchanging processing tube for multi-purpose up to 1600°C.

SPECIFICATIONS:
Power 
  • 208-240VAC, 50/60Hz, 9.0 K W
Working Temperature
  • Max 1550°C
  • Continuous: 1500°C
Recommend Heating rate
  • <10°C/min.
Processing Tube
  • Alumina Tube Size: 80 O.D x 74 I.D x 1200mm L
       
  • Ø50, Ø60, and Ø100 mm alumina tubes are available upon request. 
Heat Zone Length
  • Three heating zones:  200 mm / each (including 20 mm spacing ), 600 mm in total. 
  • Constant temperature zone:   350mm +/- 2ºC.
Temp. Profile at 1450ºC  
Overall Dimensions 860 mm × 380 mmW × 680 mmH
Temperature Controller

Temperature controller interface

The optional PC control module


Optional Eurotherm 3204 
Standard temperature controller:
  • Proportional–integral–derivative control (PID control) and auto-tune function
  • 30 segments programmed with ramping, cooling, and dwelling steps
  • Built-in over-temperature alarm and thermocouple failure alarm
  • Three S-type thermocouples.
  • +/- 1ºC temperature control accuracy
  • Default DB9 PC communication port (PC control module and software are sold separately. See optional.
Optional:  
    European Temperature Controllers (optional upgrade)
    • 1 heating channel
    • 1 program of 8 or 24 segment
    • Alarm Relay: Form 'A' and Form ‘C’ built-in 
    • Digital Communications: RS485
    • Temperature Accuracy:  ± 0.1ºC
    • Recipes: Using a PC tool recipes can be created that can be used to change the operating parameters of the 3000 simply by selecting a new recipe via the HMI. This is very useful where multiple products are processed using the same controller but require different parameters to be set
    Laptop & software (Optional)
        
    • Brand new laptop with Microsoft Windows 10 and Microsoft Office 2013 (30 days free trial) 
    • Labview Based Temperature Control System (EQ-MTS01) enables users to edit temperature profiles, manage heat-treatment recipes, record, and plot data for MTI furnaces
    • A wireless remote control provides up to 300 meters of operating range

    Thermal Block Placement
    • Two thermal blacks are included. Please picture below to learn how to place them inside a tube
    • You may order flange support to avoid tube rotating (click Pic below right)
       
    Vacuum Level
    • 10E-2 torr by a quality mechanical vacuum pump
    • 10-5 torr by molecular pump
    Heating Elements & Thermal couples
    • 24 pcs 1600ºC grade Silicon carbide Heating Elements (12 Dia. x 150 heating x  368 lengths, mm)
    • 3 pcs S type thermal couple built-in
    • Please click the picture below  to order a spare heating element
    • Notice: Customers will install SiC rod on-site by themselves as an instruction manual due to they are easy to damage during shipping
    •          
    Power Cable & Plug
    • 10 feet long 10-3 AWG heavy-duty UL-approved power cable is included. Extra length cable can be ordered by clicking picture below-left
    • The plug is not included. Browse our catalog for compatible plugs or obtain one from a local hardware store. 
                      
    Dimension & Net Weight        Weight:  160 Kg
    Operation Instructions
    Warnings

    • The tube furnaces with alumina tubes are designed for use under vacuum and low pressure < 0.02 Mpa 
    • Never use a graphite crucible in an alumina tube.
    • Attention: A two-stage pressure regulator must be installed on the gas cylinder to limit the pressure to below 3 PSI for safe operation. Click here to learn about the installation of a gas regulator.
    • The flow rate for gases should be limited to < 200 SCCM ( or 200ml/min )
    • Vacuum pressures may only be safely used up to 1500°C


    Warranty
    • One-year limited warranty with lifetime support (consumable parts such as processing tubes, and O-rings are not covered by the warranty, please order replacements at related products below.).  
    • ATTENTION: Any damages caused by the use of corrosive and acidic gases are not under the coverage of MTI One Year Limited Warranty.
    Compliance
            
    • CE Certified
    • NRTL or CSA certification is available at an extra cost.

    Related Products

    Shopping Cart
    Your cart is empty.
    Please clear the browsing history before ordering the product. Otherwise, availability and price are not guaranteed.
    Mailing Lists
    MTI sponsorships:
    MTI Sponsors Thermoelectrics Workshop

    MTI-UCSD Battery Fabrication Lab

    MTI Sponsors the Postdoctoral Awards


    Upcoming Shows:

    2024 MRS Fall Meeting December 1-6 Boston MA Booth 901

    AABC US December 9-12 Las Vegas NV Booth 812

    Click here to see our past conferences

     
    Notices
    1. Prices listed online are valid for the US market and who pay by credit card only. There will be extra charges for shipping & handling. Prices vary from country to country.
    2. We may add extra charges for paper orders and NET 30 terms due to extra labor costs.
    3. The prices listed are subject to change without notice.