Home Page
67 L/s Compact Turbo Vacuum Pump with Vacuum Gauge - Made in Germany - PV-DN63-LD
|
Transported by LTL Freight (Truck)
Product Options
Model: |
|
|
The Pfeiffer PV-DN63-LD is a compact turbo pumping station and is ideally suited for pumping down small to medium-sized volumes. The performance of the backing pump can be maximized when needed and when the pumping performance is not required, automatically reduce the pumping speed to operate in ultra-quiet mode. A precision vacuum gauge is included
Specifications:
Features
|
- Compact high-performance pumping station for all high and ultrahigh vacuum applications
- Plug and play solution with Horace 300 turbopump and multistage diaphragm pump MVP 015-4
- Intermittent mode adapts the pumping station to the process conditions and saves up to 90% of energy
- Easy integration of pressure gauges from the accessories
|
Backing Pump
|
- Model: MVP 015-4 Multistage Diaphragm Pump
- Pumping speed backing pump at 50 Hz: 0.75m^3/h | 0.44 cfm | 12.5 l.min
|
Ultimate Pressure
|
- DN 63 ISO-K: < 1 • 10^-7 mbar
- DN 63 CF-F: < 1 • 10^-8 mbar
|
Turbo Pump's Pumping Speed |
- Nitrogen N2 67 L/s
- Helium He 58 L/s
- Hydrogen H2 48 L/s
- Argon Ar 66 L/s
|
Compression Ratio |
- Nitrogen N2 > 1 * 10^11
- Helium He 1.3 * 10^7
- Hydrogen H2 1.4 * 10^5
- Argon Ar > 1 * 10^11
|
Gas throughput at full rotational speed |
- Nitrogen N2 1.3 hPa • l/s
- Helium He 2.7 hPa • l/s
- Hydrogen H2 15.3 hPa • l/s
- Argon Ar 0.54 hPa • l/s
|
Power
|
- 110 - 240 V AC, 50/60 Hz, 270W
|
Pressure gauge
|
- Pfeiffer MPT 200 AR Pirani/cold cathode combination KF25 port is included with connection cable
- 5E-9 - 1000 mbar.
|
Optional Gate Valve
|
- DN63 ISO-K or DN63 CF high vacuum bellow sealed manual gate valve (stainless steel)
|
Dimensions
|
(DN63-CF-F) (DN63-ISO-K)
|
Weight
|
38.8 lb (17.6 kg)
|
Warranty |
One-year manufacturer warranty ( The damage caused by misuse not included ) Improper use will cause all claims for liability and warranties to be forfeited. Improper use is deemed to from those mentioned below:
- Pumping of corrosive or explosive media.
- Pumping of condensing vapors.
- Pumping of liquids.
- Pumping of dust.
- Operation with improper high gas throughput.
- Operation with improper high fore-vacuum pressures.
- Operation with improper gas mode.
- Operation with improper high levels of insulated heat input.
- Venting with improper high venting rates.
- Use of the vacuum pump to generate pressure.
- The operation of the devices in areas with ionizing radiation.
- Operation in potentially explosive areas.
- The operation of the devices in systems where the turbopumps are subjected to im-pact-like stress and vibrations or the effect of periodically occurring forces.
- The use of accessories or spare parts, which are not named in this manual.
|
Operations Manual |
|
Application Notes |
- Silicon O-ring vacuum flange (for quartz tube and ceramic tube) achieves vacuum pressure up to 1E-5 torr. To obtain higher vacuum pressure (1E-6 and 1E-7torr), consider using the CF-type flange. Please check our CF flange processing tube here.
- Service for the turbopumps is recommended every 4 years under conditions with light gas loads.
|
Compliance |
- TUV Certified
- CE Approved
|
|
|
|
|
|
|
|
Sale Price: For sale with MTI Pumping Station only
| | |
|
| | |
|
Your cart is empty.
Please clear the browsing history before ordering the product. Otherwise, availability and price are not guaranteed. | |
| | |
|
MTI sponsorships: MTI Sponsors Thermoelectrics Workshop
MTI-UCSD Battery Fabrication Lab
MTI Sponsors the Postdoctoral Awards
Upcoming Shows:
|
|